| Reference |
ISO/TS 10974 ed1.0 withdrawn corrigendum |
 > preview
|
| Title |
Assessment of the safety of magnetic resonance imaging for patients with an active implantable medical device
|
| Publication date |
2012-04-23 |
Format, price (Swiss francs) and language |
 | 238.- |
| 200 pages |  | 238.- |
| 8818 Kb |
|
| |
| Abstract |
ISO/TS 10974:2012 is applicable to implantable parts of active implantable medical devices (AIMDs) intended to be used in patients who might undergo a magnetic resonance scan in 1,5T, cylindrical bore, whole body MR scanners for imaging the hydrogen nucleus.
The tests that are specified in ISO/TS 10974:2012 are type tests intended to be carried out on samples of a device to characterize interactions with the magnetic and electromagnetic fields associated with an MR scanner.
ISO/TS 10974:2012 contains test methods that are applicable to a broad class of AIMDs for the purpose of evaluating device operation against several hazards.
|
| Technical Committee |
62B - Diagnostic imaging equipment
|
| ICS Codes |
| 11.040.40 |
Implants for surgery, prosthetics and orthotics
*Including pacemakers
*Ophthalmic implants, see 11.040.70 |
|
| |
| Replaced by |
|
| |
| Work in progress |
| Project | Stage code | Forecast publication date |
|---|
| No project under development | - | - |
|