IEC 60580:2019 

Medical electrical equipment - Dose area product meters

Note: a redline version of this publication exists
IEC 60580:2019 RLV

English/French
CHF 

Do you need a multi-user copy?

English/French
CHF 

Preview

Abstract

IEC 60580:2019 is available as IEC 60580:2019 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.

IEC 60580:2019 specifies the performance and testing of Dose Area Product Meters intended to measure Dose Area Product and/or Dose Area Product Rate to which the Patient is exposed during Medical Radiological Examinations. This document is applicable to the following types of Dose Area Product Meters:
a) Field-Class Dose Area Product Meters normally used for the measurement of Dose Area Products during Medical Radiological Examinations;
b) Reference-Class Dose Area Product Meters normally used for the Calibration of Field-Class Dosimeters.
NOTE Reference-Class Dose Area Product Meters can be used as Field-Class Dose Area Product Meters.
The object of this document is
1) to establish requirements for a satisfactory level of performance for Dose Area Product Meters, and
2) to standardize the methods for the determination of compliance with this level of performance.
Two levels of performance are specified:
– a lower level of performance applying to Field-Class Dose Area Product Meters;
– a higher level of performance applying to Reference-Class Dose Area Product Meters.
IEC 60580:2019 cancels and replaces the second edition published 2000, and constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition:
a) a second class of devices is introduced with tighter uncertainty tolerances;
b) this document has been expanded to include detectors other than ionization chambers;
c) radiation qualities have been updated to the new definitions according to IEC 61267;
d) a requirement on the linearity of the dose area product rate measurement was added;
e) changed chamber light transmission requirement from 70 % to 60 %.

Look inside


Additional information

Publication typeInternational Standard
Publication date2019-11-12
Edition3.0
Available language(s)English/French
TC/SCTC 62/SC 62C - Equipment for radiotherapy, nuclear medicine and radiation dosimetryrss
ICS11.040.50 - Radiographic equipment
Stability date  2025
Pages69
File size2131 KB

The following test report forms are related:



Share your publications

Learn how to share your publications with your colleagues, using networking options.

Payment information

Our prices are in Swiss francs (CHF). We accept all major credit cards (American Express, Mastercard and Visa, JCB and CUP), PayPal and bank transfers as form of payment.


Keep in touch

Keep up to date with new publication releases and announcements with our free IEC Just Published email newsletter.

Contact customer services

Please send your enquiry by email or call us on +41 22 919 02 11 between 09:00 – 16:00 CET Monday to Friday.