ISO 80369-2:2024
Small-bore connectors for liquids and gases in healthcare applications - Part 2: Connectors for respiratory applications
Abstract
This document specifies the design and dimensions for two small-bore connectors intended to be used for connections in respiratory applications of medical devices and accessories. One connector (R1) is intended for use on medical devices and accessories subjected to pressures up to 15 kPa (e.g. a breathing system). The other connector (R2) is intended for use on medical devices and accessories subjected to higher pressures between 15 kPa and 600 kPa (e.g. oxygen therapy tubing).
NOTE 1 The pressure is related to pressure available at the source to which the medical device is connected.
NOTE 2 The intended application does not preclude the use of other connectors on medical devices or accessories within this application.
NOTE 3 Requirements for alternative connectors for this intended application are specified in ISO 80369-1.
This document does not specify requirements for the medical devices or accessories that use these connectors. Such requirements are given in device-specific standards.
NOTE 4 If a device-specific standard does not exist, the performance and material requirements specified in ISO 80369-1 can be used as guidance.
NOTE 1 The pressure is related to pressure available at the source to which the medical device is connected.
NOTE 2 The intended application does not preclude the use of other connectors on medical devices or accessories within this application.
NOTE 3 Requirements for alternative connectors for this intended application are specified in ISO 80369-1.
This document does not specify requirements for the medical devices or accessories that use these connectors. Such requirements are given in device-specific standards.
NOTE 4 If a device-specific standard does not exist, the performance and material requirements specified in ISO 80369-1 can be used as guidance.
Additional information
Publication type | International Standard |
---|---|
Publication date | 2024-07-02 |
Edition | 1.0 |
Available language(s) | English |
TC/SC | TC 62/SC 62D - Particular medical equipment, software, and systemsrss |
ICS | 11.040.10 - Anaesthetic, respiratory and reanimation equipment
11.040.20 - Transfusion, infusion and injection equipment |
Stability date | 2026 |
Pages | 50 |
File size | 2051 KB |
The following test report forms are related:
More information
Share this page
Share your publications
Learn how to share your publications with your colleagues, using networking options.
Payment information
Our prices are in Swiss francs (CHF). We accept all major credit cards (American Express, Mastercard and Visa, JCB and CUP), PayPal and bank transfers as form of payment.
Keep in touch
Keep up to date with new publication releases and announcements with our free IEC Just Published email newsletter.
Contact customer services
Please send your enquiry by email or call us on +41 22 919 02 11 between 09:00 – 16:00 CET Monday to Friday.